IBA DAP Meter
(120-IDP MICRO)
Technical Specifications
Dosimetric Performance
• Dose Area Product (DAP) Rate: Minimum 0.01 µGym2/s, Maximum 3000.00 µGym2/s
• Maximal Measurable DAP: 99999999.99 µGym2
• Minimal Dose Resolution: 0.01 µGym2
• Reproducibility: Less than 1% (at constant pressure and temperature)
• Energy Dependence: Better than +/- 8% related to 100 kV (according to IEC 60580 from 40 to 150 kV)
• Linearity: Better than +/- 2% • Measuring Time: Minimum 0.01 s
Measuring Chamber Specifications
• Active Area: 146 mm x 146 mm
• Optical Transparency: Greater than or equal to 75%
• Attenuation Equivalent: Less than 0.5 mm Al
• Electrode Spacing: 4 mm
• Chamber Voltage: 410 V +/- 5%
• Voltage Range: 15 to 28 V DC
• Mechanical Adaptation: Direct collimator mounting with adapter rails (176 mm or 167 mm distance)
• Cabling: Low voltage telecommunication or network patch cable
AC / DC Converter Specifications
• Dimensions: 95.9 mm x 55.5 mm x 30.5 mm (L x W x H)
• Input Voltage: 20 to 50 V DC / 14 to 35 V AC
• Maximum Current Consumption: 1 A
• KermaX Plus Cable Length: 650 mm
• Input Voltage Cable Length: 900 mm
• Installation Note: Not suitable for installation inside the collimator or directly into the X-ray system